Runaway is an uncommon pacemaker dysfunction, characterized by fast and erratic spikes at non-physiological rates. This infrequent but potentially lethal failure mode may be related to low battery voltage. Four single chamber pacemaker patients were analyzed (Medtronic Minix ST 8330, Minneapolis, MN, had been implanted in two patients and two CPI Triumph VR 1124, St Paul, MN, in the other two). They had been admitted because of presyncopal episodes. Typical high rate stimuli at 2000 ppm alternating with pacing at 60-65 ppm were recorded in all ECGs. Lead system tests were normal. The pulse generators had to be replaced. © 2005 Published by Elsevier Ltd on behalf of The European Society of Cardiology.

Registro:

Documento: Artículo
Título:Runaway pacemaker: A forgotten phenomenon?
Autor:Ortega, D.F.; Sammartino, M.V.; Pellegrino, G.M.M.; Barja, L.D.; Albina, G.; Segura, E.V.; Balado, R.; Laiño, R.; Giniger, A.G.
Filiación:Instituto Cardiovascular de Buenos Aires, Electrofisiología, Blanco Encalada 1543, DCO 1428 Buenos Aires, Argentina
Palabras clave:Battery depletion; Failure of capture; Runaway; aged; anamnesis; article; artificial heart pacemaker; blood pressure measurement; case report; complete heart block; electric potential; electrocardiogram; female; heart atrium flutter; heart pacing; heart rhythm; heart ventricle pressure; human; lethality; male; pacemaker electrode; pacemaker failure; priority journal; pulse generator; spike wave; syncope; Aged; Electrocardiography; Equipment Failure; Female; Humans; Male; Middle Aged; Pacemaker, Artificial
Año:2005
Volumen:7
Número:6
Página de inicio:592
Página de fin:597
DOI: http://dx.doi.org/10.1016/j.eupc.2005.06.004
Título revista:Europace
Título revista abreviado:Europace
ISSN:10995129
CODEN:EUROC
Registro:https://bibliotecadigital.exactas.uba.ar/collection/paper/document/paper_10995129_v7_n6_p592_Ortega

Referencias:

  • Furman, S., Escher, D., (1970) Principles and Techniques of Cardiac Pacing, pp. 101-102. , New York, NY: Harper and Row
  • Mickley, H., Anderson, C., Nielson, H., Runaway pacemaker: A still existing complication and therapeutic guidelines (1989) Clin. Cardiol., 12, pp. 412-414
  • Solow, E., Bacharach, B., Chung, E.K., Runaway pacemaker. Unpredictable pacemaker failure (1979) Arch. Intern. Med., 139, pp. 1190-1191
  • Carpenter, C.M., Galvin, J., Guy, M., Runaway pacemaker in an implantable cardioverter defibrillator (1998) J. Cardiovasc. Electrophysiol., 9, pp. 1008-1011
  • Katzenburg, C.A., Marcus, F., Heusinkveld, R.S., Pacemaker failure due to radiation therapy (1982) Pacing Clin. Electrophysiol., 5, pp. 156-159
  • Vijayaraman, P., Vaidya, K., Kim, S.G., Gross, J.N., Ferrick, K.J., Palma, E., Runaway pulse generator malfunction resulting from undetected battery depletion (2002) Pacing Clin. Electrophysiol., 25, pp. 220-222
  • Posse, R., Bein, I.J., D'Aiutolo, R., Arritmias cardíacas en pacientes con marcapaso electrónico implantado (1967) Rev. Arg. Cardiol., 34, pp. 47-49
  • D'Aiutolo, R., Posse, R., Tratamiento de las arritmias cardíacas (1968), pp. 69-71. , BsAs Argentina: Cabrera y Rey editores; Bohm, A., Hajdu, L., Pinter, A., Duray, G., Preda, L., Runaway syndrome and intermittent nonoutput as manifestations of end of life of a VVI pacemaker (2000) Pacing Clin. Electrophysiol., 23, pp. 2143-2144
  • Griffin, J., Smithline, H., Cook, J., Runaway pacemaker: A case report and review (2000) J. Emerg. Med., 19, pp. 177-181

Citas:

---------- APA ----------
Ortega, D.F., Sammartino, M.V., Pellegrino, G.M.M., Barja, L.D., Albina, G., Segura, E.V., Balado, R.,..., Giniger, A.G. (2005) . Runaway pacemaker: A forgotten phenomenon?. Europace, 7(6), 592-597.
http://dx.doi.org/10.1016/j.eupc.2005.06.004
---------- CHICAGO ----------
Ortega, D.F., Sammartino, M.V., Pellegrino, G.M.M., Barja, L.D., Albina, G., Segura, E.V., et al. "Runaway pacemaker: A forgotten phenomenon?" . Europace 7, no. 6 (2005) : 592-597.
http://dx.doi.org/10.1016/j.eupc.2005.06.004
---------- MLA ----------
Ortega, D.F., Sammartino, M.V., Pellegrino, G.M.M., Barja, L.D., Albina, G., Segura, E.V., et al. "Runaway pacemaker: A forgotten phenomenon?" . Europace, vol. 7, no. 6, 2005, pp. 592-597.
http://dx.doi.org/10.1016/j.eupc.2005.06.004
---------- VANCOUVER ----------
Ortega, D.F., Sammartino, M.V., Pellegrino, G.M.M., Barja, L.D., Albina, G., Segura, E.V., et al. Runaway pacemaker: A forgotten phenomenon?. Europace. 2005;7(6):592-597.
http://dx.doi.org/10.1016/j.eupc.2005.06.004