Artículo

Estamos trabajando para incorporar este artículo al repositorio
Consulte el artículo en la página del editor
Consulte la política de Acceso Abierto del editor

Abstract:

Long-term (5 years) outcome of 30 patients whose aneurysm was excluded using the Parodi endograft (PE) was assessed. Thirty patients with a mean follow-up of 59 months were the basis of the analysis. Additionally, 100 consecutive patients treated with the Vanguard Endograft (VE) were followed up for 28 months (range, 7 to 56 months). Results were analyzed and both groups compared. There were no late failures related to the loss of the integrity of the device in the PE group. Aorto aortic systems showed distal neck dilatation in high proportion of patients (80%). Aorto-uni-iliac endograft was successful in 80% of the patients. Only one late type I endoleak developed in a patient in whom the proximal stent was placed far from the renal arteries and in contact with thrombus. Also, persistent type II endoleaks were the cause of failure. No neck dilatation was noted. In the VE group, most of the failures were device related. © 2003 Elsevier Inc. All rights reserved.

Registro:

Documento: Artículo
Título:Long-term outcome after aortic endovascular repair: The Buenos Aires experience
Autor:Parodi, J.C.
Filiación:Vascular Surgery Department, Inst. Cardiovascular de Buenos Aires, Buenos Aires, Argentina
Vascular Surgery Department, Inst. Cardiovascular de Buenos Aires, Blanco Encalada 1543, 1428 Capital Federal, Argentina
Palabras clave:abdominal aorta aneurysm; artery dilatation; device; endovascular surgery; follow up; human; iliac artery; long term care; major clinical study; postoperative complication; priority journal; prosthesis failure; renal artery; review; stent; surgical technique; thrombus; treatment outcome; vascular disease
Año:2003
Volumen:16
Número:2
Página de inicio:113
Página de fin:122
DOI: http://dx.doi.org/10.1016/S0895-7967(03)00020-6
Título revista:Seminars in Vascular Surgery
Título revista abreviado:Semin. Vasc. Surg.
ISSN:08957967
CODEN:SVSUE
Registro:https://bibliotecadigital.exactas.uba.ar/collection/paper/document/paper_08957967_v16_n2_p113_Parodi

Referencias:

  • Parodi, J.C., Palmaz, J.C., Barone, H.D., Transfemoral intraluminal graft implantation for abdominal aortic aneurysms (1991) Ann Vasc Surg, 5, pp. 491-499
  • Parodi, J.C., Endoluminal stent grafts: Overview (1997) J Invasive Cardiol, 9, pp. 227-229
  • May, J., White, G.H., Waugh, R., Improved survival after endoluminal repair with second-generation prostheses compared with open repair in the treatment of abdominal aortic aneurysms: A 5-year concurrent comparison using life table method (2001) J Vasc Surg, 33 (SUPPL. 2), pp. S21-S26
  • Greenberg, R.K., Lawrence-Brown, M., Bhandari, G., An update of the Zenith endovascular graft for abdominal aortic aneurysms: Initial implantation and mid-term follow-up data (2001) J Vasc Surg, 33 (SUPPL. 2), pp. S157-S164
  • White, R.A., Clinical and design update on the development and testing of a one-piece, bifurcated, polytetra-fluoroethylene endovascular graft for abdominal aortic aneurysm exclusion: The Endologix device (2001) J Vasc Surg, 33 (SUPPL. 2), pp. S154-S156
  • Matsumura, J.S., Katzen, B.T., Hollier, L.H., Update on the bifurcated EXCLUDER endoprosthesis: Phase I results (2001) J Vasc Surg, 33 (SUPPL. 2), pp. S150-S153
  • Zarins, C.K., White, R.A., Moll, F.L., The AneuRx stent graft: Four-year results and worldwide experience 2000 (2001) J Vasc Surg, 33 (SUPPL. 2), pp. S135-S145
  • Parodi, J.C., Criado, F.J., Barone, H.D., Endoluminal aortic aneurysm repair using a balloon-expandable stent-graft device: A progress report (1994) Ann Vasc Surg, 8, pp. 523-529
  • May, J., Symposium on distortion and structural deterioration of endovascular grafts used to repair abdominal aortic aneurysms. Introduction (1999) J Endovasc Surg, 6, pp. 1-3
  • Harris, P., Brennan, J., Martin, J., Longitudinal aneurysm shrinkage following endovascular aortic aneurysm repair: A source of intermediate and late complications (1999) J Endovasc Surg, 6, pp. 11-16
  • Chaikof, E.L., Matsumura, J.S., Endovascular repair of abdominal aortic aneurysms: Problems and progress (1999) Semin Vasc Surg, 12, pp. 163-164
  • Albertini, J., Kalliafas, S., Travis, S., Anatomical risk factors for proximal perigraft endoleak and graft migration following endovascular repair of abdominal aortic aneurysms (2000) Eur J Vasc Endovasc Surg, 19, pp. 308-312
  • Walker, S.R., Macierewicz, J., Elmarasy, N.M., A prospective study to assess changes in proximal aortic neck dimensions after endovascular repair of abdominal aortic aneurysms (1999) J Vasc Surg, 29, pp. 625-630
  • Sonesson, B., Malina, M., Ivancev, K., Dilatation of the infrarenal aneurysm neck after endovascular exclusion of abdominal aortic aneurysm (1998) J Endovasc Surg, 5, pp. 195-200
  • May, J., The outcome of endoluminal repair of AAA with short proximal necks (2000) Cardiovasc Surg, 8, pp. 329-330
  • Gitlitz, D.B., Ramaswami, G., Kaplan, D., Endovascular stent grafting in the presence of aortic neck filling defects: Early clinical experience (2001) J Vasc Surg, 33, pp. 340-344
  • May, J., White, G.H., Waugh, R., Rupture of abdominal aortic aneurysms: A concurrent comparison of outcome of those occurring after endoluminal repair versus those occurring de novo (1999) Eur J Vasc Endovasc Surg, 18, pp. 344-348
  • White, R.A., Donayre, C., Wallot, I., Abdominal aortic aneurysm rupture following endoluminal graft deployment (2000) J Endovasc Ther, 7, pp. 257-262
  • Lumsden, A.B., Allen, R.C., Chaikot, E.L., Delayed rupture of aortic aneurysms following endovascular stent grafting (1995) Am J Surg, 170, pp. 174-178
  • Torsello, G.B., Klenk, E., Kasprzak, B., Rupture of abdominal aortic aneurysm previously treated by endograft stent-graft (1998) J Vasc Surg, 28, pp. 184-187
  • Umscheid, T., Stelter, W.J., Time-related alterations in shape, position, and structure of self-expanding, modular aortic stent-grafts: A 4-year single-center follow-up (1999) J Endovasc Surg, 6, pp. 17-32
  • Parodi, J.C., Berguer, R., Ferreira, L.M., Intra-aneurysmal pressure following incomplete endovascular exclusion (2001) J Vasc Surg, 34 (5), pp. 909-914
  • Lehmann, J.M., Macierewicz, J.A., Davidson, I.R., Prevention of side branch endoleaks with thrombogenic sponge: One-year follow-up (2000) J Endovasc Ther, 7, pp. 431-433
  • Van den Berg, J.C., Nolthenius, R.P., Casparie, J.W., CT-guided thrombin injection into aneurysm sac in a patient with endoleak after endovascular abdominal aortic aneurysm repair (2000) AJR Am J Roentgenol, 175, pp. 1649-1651
  • Makaroun, M.S., Deaton, D.H., Is proximal aortic neck dilatation after endovascular aneurysm exclusion a cause for concern? (2001) J Vasc Surg, 33 (SUPPL. 2), pp. S39-S45
  • Albertini, J., Kalliafas, S., Travis, S., Anatomical risk factors for proximal perigraft endoleak and graft migration following endovascular repair of abdominal aortic aneurysms (2000) Eur J Vasc Endovasc Surg, 19, pp. 308-312
  • Harris, P.L., Vallabhaneni, S.R., Desgranges, P., Incidence and risk factors of late rupture, conversion, and death after endovascular repair of infrarenal aortic aneurysms: The EUROSTAR experience (2000) J Vasc Surg, 32, pp. 739-749. , European Collaborators on Stent/graft techniques for aortic aneurysm repair
  • Broeders, I.A., Blankensteijn, J.D., Wever, J.J., Mid-term fixation stability of the EndoVascular Technologies endograft (1999) Eur J Vasc Endovasc Surg, 18, pp. 300-307. , EVT Investigators
  • Ohki, T., Veith, F.J., Patient selection for endovascular repair of abdominal aortic aneurysms: Changing the threshold for intervention (1999) Semin Vasc Surg, 12, pp. 226-234
  • Armon, M.P., Wenham, P.W., Hopkinson, B.R., Suitability for endovascular aneurysm repair in an unselected population (2001) Br J Surg, 88, pp. 889-890. , Br J Surg 88:77-81, 2001

Citas:

---------- APA ----------
(2003) . Long-term outcome after aortic endovascular repair: The Buenos Aires experience. Seminars in Vascular Surgery, 16(2), 113-122.
http://dx.doi.org/10.1016/S0895-7967(03)00020-6
---------- CHICAGO ----------
Parodi, J.C. "Long-term outcome after aortic endovascular repair: The Buenos Aires experience" . Seminars in Vascular Surgery 16, no. 2 (2003) : 113-122.
http://dx.doi.org/10.1016/S0895-7967(03)00020-6
---------- MLA ----------
Parodi, J.C. "Long-term outcome after aortic endovascular repair: The Buenos Aires experience" . Seminars in Vascular Surgery, vol. 16, no. 2, 2003, pp. 113-122.
http://dx.doi.org/10.1016/S0895-7967(03)00020-6
---------- VANCOUVER ----------
Parodi, J.C. Long-term outcome after aortic endovascular repair: The Buenos Aires experience. Semin. Vasc. Surg. 2003;16(2):113-122.
http://dx.doi.org/10.1016/S0895-7967(03)00020-6