Abstract:
A method is proposed for wavelength calibration of the reversed-optics polychromators of photodiode array spectrophotometers. The central wavelength of each diode is determined using classical line emission sources and an appropriate algorithm. The method can be used even when the light from the slit image formed on the detector falls on several diodes, whether due to a large slit to diode ratio, detector window effects or other causes.
Referencias:
- (1987) Standard Method for Describing and Measuring Performance of Ultraviolet, Visible and Near Infrared Spectrophotometen, Annual Book of ASTM Standards, p. 408. , vol. 14.01, 1987, E 275–83, American Society for Testing and Materials (ASTM), Philadelphia, PA
- (1981) Standards in Absorption Spectrometry, p. 130. , Burgess, C., and Knowles, A., Chapman and Hall, London and New York
- Mavrodineanu, R., (1975) NBS Spec. Publ. 260-51, pp. 54-64. , Washington, DC
- (1987) Advances in Standards and Methodology in Spectrophotometry, , Burgess, C., and Mielenz, K. D., Elsevier, Amsterdam, pp. 99, 125 and 303
- (1987) Standard Practice for the Periodic Calibration of Narrow Band-Pass Spectrophotometers, Annual Book of ASTM Standards, p. 824. , vol. 14.01, 1987, E 925–83, American Society for Testing and Materials (ASTM), Philadelphia, PA
- McNeirney, J., Slavin, M., (1962) Appl. Opt., 1, p. 365
- (1984) Practical Absorption Spectrometry, p. 221. , Knowles, A., and Burgess, C., Chapman and Hall, London and New York
- Burgess, C., (1977) UV Spectrom. Group Bull., 5, p. 77
- Fog, J., Osnes, E., (1962) Analyst, 87, p. 760
Citas:
---------- APA ----------
Berlot, P.E. & Locascio, G.A.
(1991)
. Ultraviolet-visible photodiode array spectrophotometer wavelength calibration method. A practical computer algorithm. The Analyst, 116(3), 313-316.
http://dx.doi.org/10.1039/AN9911600313---------- CHICAGO ----------
Berlot, P.E., Locascio, G.A.
"Ultraviolet-visible photodiode array spectrophotometer wavelength calibration method. A practical computer algorithm"
. The Analyst 116, no. 3
(1991) : 313-316.
http://dx.doi.org/10.1039/AN9911600313---------- MLA ----------
Berlot, P.E., Locascio, G.A.
"Ultraviolet-visible photodiode array spectrophotometer wavelength calibration method. A practical computer algorithm"
. The Analyst, vol. 116, no. 3, 1991, pp. 313-316.
http://dx.doi.org/10.1039/AN9911600313---------- VANCOUVER ----------
Berlot, P.E., Locascio, G.A. Ultraviolet-visible photodiode array spectrophotometer wavelength calibration method. A practical computer algorithm. 1991;116(3):313-316.
http://dx.doi.org/10.1039/AN9911600313